• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC SAFE-T-CENTESIS TRAY 6FR X 16CM; THORACENTESIS TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC SAFE-T-CENTESIS TRAY 6FR X 16CM; THORACENTESIS TRAY Back to Search Results
Catalog Number PIG1260T
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 03/17/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, images were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 09/2026).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported by customer that the safe t centesis kits- lawson 218295 have started containing a different vial of lidocaine.The vial has shattered when opening on multiple occasions causing lacerations on 2 employees and cutting gloves.The kits are also rendered unusable when this happens because of glass across the field.Quote from customer- the saf t centesis kits- lawson 218295 have started containing a different vial of lidocaine.The vial has shattered when opening on multiple occasions causing lacerations on 2 employees and cutting gloves.The kits are also rendered unusable when this happens because of glass across the field.Saf-t- centesis kit lot: 0001589013.
 
Additional Manufacturer Narrative
H10: manufacturing review: a review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections related to this incident were found.Product undergoes inspections during manufacturing, no issues related to the reported incident were identified, all procedural and functional requirements for product release have been met.H10: investigation summary: four photos were provided for analysis.From the provided photos we are unable to confirm damage.A root cause for the failure could not be confirmed.H10: b5, d4, g4, h2 (correction and additional information), h6 (annex g).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported by customer that the safe t centesis kits have started containing a different vial of lidocaine.The vial has shattered when opening on multiple occasions causing lacerations on 2 employees and cutting gloves.The kits are also rendered unusable when this happens because of glass across the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE-T-CENTESIS TRAY 6FR X 16CM
Common Device Name
THORACENTESIS TRAY
Manufacturer (Section D)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer (Section G)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer Contact
brett curtice
75 n. fairview drive
vernon hills, IL 60061
8015652341
MDR Report Key21773893
Report Number1625685-2025-00031
Device Sequence Number12820944
Product Code PXI
UDI-Device Identifier10885403108365
UDI-Public(01)10885403108365(11)240911(17)260930(10)0001589013
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Catalogue NumberPIG1260T
Device Lot Number0001589013
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2025
Supplement Date Received by Manufacturer05/22/2025
Initial Report FDA Received Date04/04/2025
Supplement Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-