| Model Number |
102812-81A |
| Medical Device Problem Code |
Premature End-of-Life Indicator (1480)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
|
| |
|
Event or Problem Description
|
|
On march 7, 2025, senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
|
| |
|
Additional Manufacturer Narrative
|
|
The sensor replacement alert was first presented to the user on (b)(6) 2025, day 143 after insertion.The sensor was tested in-house, and the review of investigation analysis revealed a loss of chemical performance.The system correctly disabled the sensor when it detected the performance failure, and the system's self-test functions were working normally.The root cause of the performance failure was due to the sensor's hydrogel oxidation.As part of resolution, a return material authorization was issued for sensor replacement.No further resolution was necessary for this complaint.B4.Date of this report 04 june 2025.G3.Date received by the manufacturer? 04 june 2025.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 3231.H6.Investigation conclusions updated to 4307.
|
| |
|