| Model Number |
ACU0T0 |
| Medical Device Problem Codes |
Device Difficult to Setup or Prepare (1487); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that during intraocular lens (iol) priming, lens stuck in delivery system and trailing haptic folded wrong.Lens did not come out, it was blocked.The procedure was completed on same day.There was no patient contact.Additional information has been requested.There are three medical device reports associated with this report.This file is 1 of 3.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.A photo was provided showing the carton endcap.Product history records were reviewed, and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The product investigation could not identify a root cause.The product has not been received to evaluate.Lens may become stuck in the device: if inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to become ¿stuck¿ in the device if the operating room temperature is too high (more than 23 degree celsius / 73-degrees fahrenheit) lens folding consistency is negatively affected, the lens is more adherent, and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to requests for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|