EXACTECH, INC. TRULIANT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-020-11-0220 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitants: (b)(6) 02-020-11-0220 truliant femoral component.The revision reported was likely the result of a combination of aseptic loosening between the femoral and bone and prosthesis wear.Femoral loosening may have led to an increase in cement and bone particles in the joint space, component malalignment and resulted in prosthesis wear or prosthesis wear may have contributed to particle-induced osteolysis and subsequent loosening.Additionally, a contributing factor to the reported wear may have been the insert being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the sequence of events could not be assessed as the devices were not available for evaluation.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
As reported, the patient had an initial left tka on (b)(6) 2010.The patient had a recalled implant and was revised on (b)(6) 2023.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.This is 1 of 2 reports for this event.See mdr # 1038671-2023-02850.
|
| |
|