Model Number SCORPION NEEDLE, KNEE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a meniscal repair surgery the needle was stuck in the scorpion.The needle broke inside the scorpion.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause the reported failure can be attributed to user error of the device due to excessive force being used during firing of the needle thus, breaking the needle and causing a blockage within the shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCORPION NEEDLE, KNEE
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21777754
Report Number1220246-2025-01249
Device Sequence Number12710186
Product Code LXH
UDI-Device Identifier00888867266520
UDI-Public(01)00888867266520(11)240411(17)290331(10)15233412
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSCORPION NEEDLE, KNEE
Device Catalogue NumberAR-12990N-1
Device Lot Number15233412
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2025
Supplement Date Received by Manufacturer08/11/2025
Initial Report FDA Received Date04/04/2025
Supplement Report FDA Received Date08/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-