EXACTECH, INC. TRAPEZOID TIBIAL TRAY SZ 4F/4T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
204-04-44 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Osteolysis (2377)
|
| Date of Event |
12/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: ps tibial insert size 4, 9mm (cat#204-24-09 / serial # (b)(6).Three peg patella 32mm (cat# 200-02-32/ serial # (b)(6).The revision reported was likely due to a combination of osteolysis and loosening between the patellar and tibial components and the underlying bone over the course of 11 years of implantation.Patellar maltracking may have been a contributing factor to the patellar wear and/or loosening.Unintended contact between the trochlear ridge of the femoral component during knee extension and/or third body wear may have been a contributing factor to the tibial insert damage.Patellar and tibial loosening cannot be confirmed as the devices were not available for evaluation and limited clinical information was provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
As reported, approximately ten years post initial right tka, the 70 y/o male patient had a revision due to loose patella, poly issues and osteolysis of the tibia.Surgeon changed the liner, patella and tibial tray and also used an extension stem.There were no parts or pieces fell into the patient wound site.Patient was last known to be in stable condition following the event.X-rays and photos received.The devices are not available for evaluation as they were discarded by the hospital.No other patient information/medical history reported.
|
| |
|
Search Alerts/Recalls
|
|
|