MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3542650 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
12/30/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient who underwent primary breast augmentation with a mentor siltex round moderate profile 325cc saline prosthesis experienced left sided deflation post procedure.As a result, explant without replacement was performed on (b)(6) 2025.The patient is fully recovered with no residual effects.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation review could be performed.Reason for device explant and/or reoperation: deflation.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On april 9, 2025, mentor became aware that it erroneously stated that the deflation reported was left sided.The deflation was right sided.In addition to the right sided deflation, the patient also experienced left sided baker grade ii capsular contracture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The impacted product was received by the failure analysis lab on may 2, 2025.The investigation of the returned device was completed by the failure analysis lab on may 9, 2025.Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned device revealed that the breast implant was found to have a tear on the posterior view measuring approximately 0.6 cm.Leak testing was performed, in accordance with mentor procedures, and no additional leak sites were detected during the analysis.Microscopic examination was performed, and the cause of the deflation could not be identified.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of saline implants include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|