• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 3542650
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 12/30/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient who underwent primary breast augmentation with a mentor siltex round moderate profile 325cc saline prosthesis experienced left sided deflation post procedure.As a result, explant without replacement was performed on (b)(6) 2025.The patient is fully recovered with no residual effects.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation review could be performed.Reason for device explant and/or reoperation: deflation.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On april 9, 2025, mentor became aware that it erroneously stated that the deflation reported was left sided.The deflation was right sided.In addition to the right sided deflation, the patient also experienced left sided baker grade ii capsular contracture.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The impacted product was received by the failure analysis lab on may 2, 2025.The investigation of the returned device was completed by the failure analysis lab on may 9, 2025.Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned device revealed that the breast implant was found to have a tear on the posterior view measuring approximately 0.6 cm.Leak testing was performed, in accordance with mentor procedures, and no additional leak sites were detected during the analysis.Microscopic examination was performed, and the cause of the deflation could not be identified.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of saline implants include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SILTEX ROUND MODERATE PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key21778243
Report Number1645337-2025-03682
Device Sequence Number12729779
Product Code FWM
UDI-Device Identifier00081317001799
UDI-Public00081317001799
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2002
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3542650
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/14/2025
Supplement Date Received by Manufacturer04/09/2025
05/02/2025
Initial Report FDA Received Date04/04/2025
Supplement Report FDA Received Date04/23/2025
05/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-