Model Number 105-7100-060
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report that a procedure went as planned.However, when the microcatheter was removed at the end of the procedure, the distal end snapped and about 5cm was left behind in the patient.An attempt was made to remove the catheter with a snare, but it was unsuccessful.There was no friction or difficulty during injection.There was no force applied during delivery or removal.The catheter tip was entrapped/stuck.There was no vasospasm.The catheter was not stuck inside the guide catheter.No surgical or medicinal intervention was required.The broken segments remain in the patient in the imax.The reported device and any accessory devices were prepared and the catheter was flushed as indicated in the instructions for use (ifu).No patient symptoms or complications were associated with this event.The patient was undergoing surgery for onyx embolization of mma.It was noted the patient's vessel tortuosity was normal.The access vessel was the external carotid artery (eca) - internal maxillary artery (imax) - middle meningeal artery (mma) with a 2.5mm diameter.Ancillary devices include an onyx 18 liquid embolic.Additional information was received that the procedure was completed with an attempt to snare out remaining the microcatheter, but about 15cm was left in patient.It was clarified that the catheter was broken.It was believed that the catheter was entrapped in onyx.The cause of the break and entrapment was not determined.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONYX AVM
Common Device Name
AGENT, INJECTABLE, EMBOLIC
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key21778677
Report Number2029214-2025-00889
Device Sequence Number21040219
Product Code MFE
UDI-Device Identifier00847536006171
UDI-Public00847536006171
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number105-7100-060
Device Catalogue Number105-7100-060
Device Lot NumberB813025
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2025
Initial Report FDA Received Date04/04/2025
Date Device Manufactured11/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age89 YR
Patient SexMale
-
-