Additional information is provided in sections d.9, h.3, h.6 and h.11.A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A review of the database for the non-conformance events and the complaints showed that the given lot number was not concerned by any deviation and that there are no additional complaints associated with it.The investigation is concluded based on the sample evaluation outlined below.The returned instrument was received at the investigation site in its inner blister and covered with the tyvek lid foil showing the lot information.The forceps was returned in a close state, and it shows clear signs of surgery residues.The returned instrument was visually inspected with the aid of a photomicroscope using various magnifications.The visual inspection revealed no immediate issues such as bumps or deformation of the forceps arms.The device was then functionally tested, it was noticed that no actuation of the forceps is possible.The actuation mechanism itself is still in working condition, but the forceps stuck in a close position.The forceps can only get opened again by manually pushing down the metal shaft, indicating that the bonding connection between the shaft and stiffening sleeve got broken.As the blister was opened by the customer, the product was handled.Therefore, the point in time when the malfunction occurred can no longer be determined, nor can the strain put on the device be reconstructed.The customer¿s complaint is confirmed, but a root cause for the connection failure cannot be determined.It cannot be determined how or where the damage to the sample occurred.Therefore, the root cause for the customer's reported event could not be determined.The exact root cause for the customers reported event is unknown, therefore, specific action cannot be taken.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed associated with the reported product and event.No action has been identified for this reported event.The manufacturer internal reference number is: (b)(4).
|