• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC. Biozorb; MARKER RADIOGRAPHIC IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC. Biozorb; MARKER RADIOGRAPHIC IMPLANTABLE Back to Search Results
Model Number F0304
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abscess (1690); Bacterial Infection (1735); Breast Discomfort/Pain (4504); Swelling/ Edema (4577)
Date of Event 12/09/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Hologic is submitting this report in accordance with 21.Cfr part 803.The submission is based upon the currently available information.The submission of this report does not represent a confirmation of device malfunction involving the hologic products in the event described.Lot number of the device not provided by the complainant; therefore, the udi, expiration and manufacturing dates are not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event or Problem Description
It was reported that a patient received a biozorb device on (b)(6) 2020, patient reported that she experienced pain, swelling and infection at the site of implantation.Patient underwent excisional biopsy on her left breast on (b)(6) 2023 and the biozorb device was removed then.No other information available.
 
Additional Manufacturer Narrative
An investigation was conducted: brief description: it was reported that a patient received a biozorb device on (b)(6) 2020, patient reported that she experienced pain, swelling and infection at the site of implantation.Patient underwent excisional biopsy on her left breast on (b)(6) 2023, and the biozorb device was removed then.No other information available.Investigation methods and findings: the sample was not returned to the post market quality laboratory for further investigation; therefore, the inspection according to the biozorb post market instructions was not able to be conducted for this complaint.Additionally, there was limited patient-related information provided for this event; consequently, a comprehensive investigation could not be conducted.The harms identified, based on clinical symptoms and hazardous situations for this complaint are: - pain, serious (pain, sleep dysfunction, device palpability, failure to resorb).- infection, serious (infection).- surgery/device explantation, calcifications).- inflammation, serious (limited mobility, lymphedema, swelling).Cause/root cause: the described issue was investigated through a root cause analysis conducted under capa.The initial purpose of this capa was to evaluate the biozorb product for a root cause linking the product design to the reported complaint issues.The results of the root cause assessment did not find any direct link between the device and the reported complaint issues.Consequently, the implementation activities from capa were considered no longer applicable since hologic has ceased selling the biozorb product with no intention of returning it to market.In addition, capa is currently evaluating the biozorb product¿s entire dhf and is actively under remediation.Furthermore, a voluntary recall for the biozorb product was initiated on (b)(6) 2024.Conclusion: the reported issue could not be confirmed as no sample was returned for evaluation.The risk trending analysis does not increase the risk zone index.The potential root cause analysis was assessed under capa.Capa is currently evaluating the biozorb product¿s entire dhf and is actively under remediation.The risks associated with this event are further detailed and evaluated against the product and its hazard analysis within the health risk assessment biozorb, complaint evaluation.Future events will be monitored and trended.Complaint confirmed: no.Device history record (dhr) review: not performed.No lot number was provided; therefore, no dhr could be located or reviewed.
 
Event or Problem Description
Hologic has received correspondence arising from litigation.The additional medical information provided, revealed the f0304 catalog number of the device and the lot number d1-190305 of the device.On (b)(6) 2022 the patient had the left breast abscess aspirated (which was located at the site of the previous lumpectomy).On (b)(6)2023 the patient presented with pain, swelling, spontaneous drainage and abscess at the implant site.The excision biopsy on left breast and the biozorb device explantation were performed on the same date.No other relevant information was provided.This report is being submitted pursuant to 21 cfr part 803 based on information made available to hologic through litigation.Consistent with 21 cfr 803.16, the submission of this report is not intended, and shall not be construed, as an admission, acknowledgment, or confirmation by hologic that the device caused or contributed to the reportable event.
 
Additional Manufacturer Narrative
Hologic has received correspondence arising from litigation.The additional medical information provided, revealed the f0304 catalog number of the device and the lot number d1-190305 of the device.On (b)(6) 2022 the patient had the left breast abscess aspirated (which was located at the site of the previous lumpectomy).On (b)(6) 2023 the patient presented with pain, swelling, spontaneous drainage and abscess at the implant site.The excision biopsy on left breast and the biozorb device explantation were performed on the same date.No other relevant information was provided.This report is being submitted pursuant to 21 cfr part 803 based on information made available to hologic through litigation.Consistent with 21 cfr 803.16, the submission of this report is not intended, and shall not be construed, as an admission, acknowledgment, or confirmation by hologic that the device caused or contributed to the reportable event.A device history record (dhr) review was conducted for the reported lot number.The device was released meeting all qa specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Biozorb
Common Device Name
MARKER RADIOGRAPHIC IMPLANTABLE
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free park zone b24
alajuela 20102
MDR Report Key21786093
Report Number1222780-2025-00164
Device Sequence Number15633592
Product Code NEU
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
NOTAVAILABLE
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Model NumberF0304
Device Catalogue NumberF0304
Device Lot NumberD1-190305
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2025
Supplement Date Received by ManufacturerNot provided
04/10/2026
06/01/2026
09/23/2026
Initial Report FDA Received Date04/07/2025
Supplement Report FDA Received Date04/07/2025
04/14/2026
06/16/2026
09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-