Correction to d4(gtin) the complaint could be confirmed, since the information for evaluation matches the alleged failure.Upon further investigation of the ct scans by healthcare professionals the following was observed: the ct scan shows clear loosening of the tibial component with some radiolucence and larger cysts.The implantation was in jan.2024.The bone shows some reaction which could potentially be a sign of infection.However, the case is described as aseptic loosening and the surgeon writes, that he suspects missing integration in the bone as the cause of the failure.From the images and the clinical facts so far the underlying cause remains unclear.It could be patient related due to poor bone substance.Technical problems with the implantation could also be the reason for the failure or ¿there is a systematic issue with lack of bony integration with the tibia¿, which then would have to be regarded as a device related failure.There is, however, no specific sign for any of these reasons and none of them can be excluded with the provided information.However, failure of total ankle replacement tar over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient radiological and clinical information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information e.G.Clinical status, exact symptoms and range of motion of the patient, assessment of the treating physician is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and or the device in relation to cause of the failure.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|