BOSTON SCIENTIFIC CORPORATION SYNERGY XD MONORAIL EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; CORONARY DRUG-ELUTING STENT Back to Search Results
Catalog Number H7493941824270
Health Effect - Clinical Codes Cardiac Arrest (1762); Chest Pain (1776); Discomfort (2330); Respiratory Arrest (4461)
Date of Event 03/24/2025
Type of Reportable Event Death
Event or Problem Description
Patient arrived in cath lab for a cardiac catheterization.Procedure began with access to the right groin and diagnostic imaging showed a lesion in the proximal lad.Lesion was predilated with a 2.0 x 20mm takeru balloon.A 2.75 x 24mm synergy des was then passed down and across the lesion.Stent deployed to 20 atms.Upon attempted removal of delivery catheter, resistance was met.Catheter then released and was removed where it was noted the catheter was broken.Angiography showed the balloon markers were still in the vessel.Patient began to complain of chest discomfort and patient was intubated.Vascular surgeon was called to assist and successfully retrieved broken piece from body.Patient distress continued to escalate.Cardiac and respiratory arrest shortly followed.Cpr began and attempts to place an impella device unsuccessful.Patient expired in cath lab.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD MONORAIL EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
MDR Report Key21797169
Report NumberMW5168702
Device Sequence Number8210326
Product Code NIQ
UDI-Device Identifier08714729980902
UDI-Public(01)08714729980902(17)260415(10)33881804
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Voluntary
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 04/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/15/2026
Device Catalogue NumberH7493941824270
Device Lot Number33881804
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/08/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age56 YR
Patient SexMale
Patient Weight124 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-