• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC EPIQ 5; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC EPIQ 5; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Catalog Number 795204
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
As part of capa 5677695, it was identified that retrospective review activities were warranted for intracavity transducer damage if physically split, fractured, or contains sharp edges on the transducer, regardless of whether the device is in clinical use.Philips ultrasound is reporting this complaint as part of the retrospective review completed for intracavity transducer damage.It was reported the c10-3v transducer had physical damage to the lens causing it to fail the electrical leakage test.There was no patient or user harm associated with this event.The service engineer confirmed the reported issue and replaced the transducer to resolve the customer¿s immediate concerns.Philips has started an investigation, and upon completion, a follow up report will be submitted.
 
Additional Manufacturer Narrative
H11: the initial report inadvertently reported an incorrect date received by mfg (g3) and should have been 21mar2025.A thorough investigation was unable to be performed to determine the cause of the reported problem.The transducer was no longer available; therefore, no repair or evaluation data could be obtained.No additional investigation is possible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ 5
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
roxanne ramirez
22100 bothell everett highway
bothell, WA 98021
MDR Report Key21800705
Report Number3019216-2025-000076
Device Sequence Number8311797
Product Code IYN
UDI-Device Identifier00884838097896
UDI-Public(01)00884838097896
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number795204
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2022
Supplement Date Received by Manufacturer07/07/2025
Initial Report FDA Received Date04/08/2025
Supplement Report FDA Received Date07/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-