| Catalog Number |
CU-25122-F |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, "on (b)(6) 2025, hemodialysis unit, (b)(6) hospital the doctor gave feedback the swg has a quality issue during use on the patient." the user changed to a new set to resolve the issue.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The actual device was not returned; however, the customer provided two photos for analysis.The complaint of a kinked guidewire was able to be confirmed by the photos, which show a guidewire in the clinical setting with obvious kinks and bends, however, a complete visual inspection to determine the cause of the damage could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested." without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported, "on (b)(6) 2025, hemodialysis unit, (b)(6) hospital the doctor gave feedback the swg has a quality issue during use on the patient." the user changed to a new set to resolve the issue.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
|
| |
|
Search Alerts/Recalls
|