| Model Number |
4554 |
| Medical Device Problem Codes |
Defective Device (2588); Material Integrity Problem (2978)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inappropriate Device Stimulation of Tissue (4928)
|
| Date of Event |
03/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this cardiac resynchronization therapy pacemaker (crt-p) experienced sensation near the device and down the lead.Touching and pressing the area feels like an electrical shock.Boston scientific technical services (ts) discussed that it was a possible insulation exposure resulting in stimulation.The whole system was explanted due to infection.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report was amended to update international medical device regulators forum (imdrf) codes.Incomplete coding of this event was identified during a retrospective review of complaints associated with capa-8580, which focused on identifying and remediating incomplete coding of events.Added patient code (b)(6).A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of damaged/defective was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.The device was not returned for analysis; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with this cardiac resynchronization therapy pacemaker (crt-p) experienced sensation near the device and down the lead.Touching and pressing the area feels like an electrical shock.Boston scientific technical services (ts) discussed that it was a possible insulation exposure resulting in stimulation.The whole system was explanted due to infection.No additional adverse patient effects were reported.
|
| |
|