BOSTON SCIENTIFIC CORPORATION ACUITY? STEERABLE; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 4554
Medical Device Problem Codes Defective Device (2588); Material Integrity Problem (2978)
Health Effect - Clinical Codes Unspecified Infection (1930); Inappropriate Device Stimulation of Tissue (4928)
Date of Event 03/11/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this cardiac resynchronization therapy pacemaker (crt-p) experienced sensation near the device and down the lead.Touching and pressing the area feels like an electrical shock.Boston scientific technical services (ts) discussed that it was a possible insulation exposure resulting in stimulation.The whole system was explanted due to infection.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This report was amended to update international medical device regulators forum (imdrf) codes.Incomplete coding of this event was identified during a retrospective review of complaints associated with capa-8580, which focused on identifying and remediating incomplete coding of events.Added patient code (b)(6).A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of damaged/defective was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.The device was not returned for analysis; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
 
Event or Problem Description
It was reported that the patient with this cardiac resynchronization therapy pacemaker (crt-p) experienced sensation near the device and down the lead.Touching and pressing the area feels like an electrical shock.Boston scientific technical services (ts) discussed that it was a possible insulation exposure resulting in stimulation.The whole system was explanted due to infection.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUITY? STEERABLE
Common Device Name
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21806153
Report Number2124215-2025-21987
Device Sequence Number8834670
Product Code NVY
UDI-Device Identifier00802526419324
UDI-Public(01)00802526419324(17)110423
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P050046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/23/2011
Device Model Number4554
Device Catalogue Number4554
Device Lot Number175488
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2025
Supplement Date Received by Manufacturer04/21/2026
Initial Report FDA Received Date04/09/2025
Supplement Report FDA Received Date04/23/2026
Date Device Manufactured04/23/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age75 YR
Patient SexMale
-
-