| Model Number |
IPN920517 |
| Medical Device Problem Code |
Material Disintegration (1177)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "green silicone joint was non-permeable.Device was changed." additional information received reports "when the seal was inserted, the device disintegrated in a circular fashion".There was no clinical consequences for the patient.No medical intervention required.The patient's current condition is reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The customer returned one compression sleeve and lidstock from a hemodialysis kit for analysis.Signs of use were observed.The connector assembly was not returned.Visual analysis revealed damage to one end of the compression sleeve.The sleeve appeared to be cut in a spiral pattern.Additionally, thread indentations were visible along the cut edge.The damage to the compression sleeve is consistent with the compression sleeve being misplaced within the compressing fitting/cap which would cause the metal cannulas on the connector assembly to catch onto the edge of the compressing sleeve as the components are assembled.When the compression fitting/cap and sleeve are twisted onto the juncture hub, the compression sleeve can get damaged/caught in the threads.The inner diameter of the compression sleeve measured 0.210" which is within the specification limits of 0.207"-0.213" per the extrusion product drawing.The outer diameter of the compression sleeve measured 0.2595" which is within the specification limits of 0.257"-0.263" per the extrusion product drawing.The length of the compression sleeve measured 0.3575" which is within the specification limits of 0.34"-0.36" per the extrusion product drawing.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "warning: green compression sleeve must be present when threading compression cap onto hub connection assembly.Failure to do so may result in air embolism, blood loss, or catheter separation.Precaution: ensure compression sleeve is securely positioned inside threaded compression cap.Avoid attempts to place compression sleeve onto hub connection assembly cannula and then apply threaded compression cap.Incomplete compression may occur causing catheter separation." the complaint of damage to the compression sleeve was able to be confirmed through complaint investigation of the returned sample.The compression sleeve was damaged/torn upon return which is consistent with misalignment of the compression sleeve within the compression fitting/cap prior to twisting on the connector assembly.Based on the customer report and sample received, unintentional use error likely caused or contributed to the reported event.Teleflex will continue to monitor and trend on complaints of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that "green silicone joint was non-permeable.Device was changed." additional information received reports "when the seal was inserted, the device disintegrated in a circular fashion".There was no clinical consequences for the patient.No medical intervention required.The patient's current condition is reported as unknown.
|
| |
|
Search Alerts/Recalls
|