| Model Number |
0663 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
03/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead was part of a system revision due to infection.There was no additional adverse patient effects were reported.This rv lead was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, additional information from product investigation indicates that the allegation against this lead was not confirmed.This lead was analyzed, passed all the baseline tests and exhibited normal lead functions.This supplemental is being filed to capture d9: returned to manufacturer date and investigation results.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead was part of a system revision due to infection.There was no additional adverse patient effects were reported.This rv lead was explanted.
|
| |
|