• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW FIBEROPTIX ULTRA 8 IAB: 8FR 30CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW FIBEROPTIX ULTRA 8 IAB: 8FR 30CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAB-05830-LWS-SC
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported "potential iab abrasion- persistent helium loss 2 alarms unresolved".Additional information states that the iab catheter was removed and replaced.No patient injury or consequence reported.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4) the reported complaint for "persistent helium loss 2 alarms" is confirmed based on the customer provided photo with the complaint report.In the photo, the "possible helium loss 2" alarm was noted on the iabp display screen.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the helium loss alarm.The root cause of the complaint is undetermined.No further action is required at this time.The reported complaint will be monitored for any developing trends.If the product is returned at a later date, a full investigation of the sample will be completed.
 
Event or Problem Description
It was reported "potential iab abrasion- persistent helium loss 2 alarms unresolved".Additional information states that the iab catheter was removed and replaced.No patient injury or consequence reported.The patient's current condition is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW FIBEROPTIX ULTRA 8 IAB: 8FR 30CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
sachin sony
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key21808265
Report Number3010532612-2025-00285
Device Sequence Number18394688
Product Code DSP
UDI-Device Identifier00801902034724
UDI-Public(01)00801902034724(11)250210(17)270131(10)18F25A0023(21)
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K021462
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberIAB-05830-LWS-SC
Device Lot Number18F25A0023
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2025
Supplement Date Received by Manufacturer04/30/2025
Initial Report FDA Received Date04/09/2025
Supplement Report FDA Received Date05/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
-
-