| Catalog Number |
IAB-05830-LWS-SC |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "potential iab abrasion- persistent helium loss 2 alarms unresolved".Additional information states that the iab catheter was removed and replaced.No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) the reported complaint for "persistent helium loss 2 alarms" is confirmed based on the customer provided photo with the complaint report.In the photo, the "possible helium loss 2" alarm was noted on the iabp display screen.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the helium loss alarm.The root cause of the complaint is undetermined.No further action is required at this time.The reported complaint will be monitored for any developing trends.If the product is returned at a later date, a full investigation of the sample will be completed.
|
| |
|
Event or Problem Description
|
|
It was reported "potential iab abrasion- persistent helium loss 2 alarms unresolved".Additional information states that the iab catheter was removed and replaced.No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|