| Catalog Number |
M0036180840 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This is 1/2 reports.
|
| |
|
Event or Problem Description
|
|
It was reported during the embolization procedure, the subject coil was detached while advancing inside the catheter.The detached coil was removed using the guidewire.While removing the subject coil, another coil that was already implanted, came out with the detached coil.The procedure was completed successfully with no clinical consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to the automated mes/dhr (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual and functional testing as well as physical analysis cannot be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.It was reported that the coil was detached inside the microcatheter while advancing.While removing the coil, it pulled out a previously detached coil.The device was confirmed to be in good condition prior to use.There is no indication of a use error.An assignable cause of undeterminable will be assigned to this complaint as the product was not returned and review and analysis of all available information fails to indicate an assignable cause or probable assignable cause.
|
| |
|
Event or Problem Description
|
|
It was reported during the embolization procedure, the subject coil was detached while advancing inside the catheter.The detached coil was removed using the guidewire.While removing the subject coil, another coil that was already implanted, came out with the detached coil.The procedure was completed successfully with no clinical consequences to the patient.
|
| |
|