Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported rupture.This record is for the right side.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaint codes are: - rupture: observed a broken assessed as a surgical impact opening.(shell thickness within spec).As per the investigation procedure, non-penetrating nick were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.Additional, changed, and/or corrected data: d9; h3; h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device has been explanted and replaced.
|
| |
|