| Catalog Number |
72202745 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
03/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an meniscus repair surgery, the biosure screw deformed and broke.The procedure was successfully completed using a smith and nephew back up device.It is unknown if there was a delay.All the pieces were removed form the patient using tweezers.The back up was used in an additional drilled hole and there was a void left in the patient.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: the reported device was received for evaluation.A visual inspection revealed the device was returned in original packaging with the batch number of the complaint on the label.One fin is fractured from the device and was not returned.A second fin is severely bent and deformed.The third fin is bent and the fourth has no visible defect.Based on the condition of the product material found during visual inspection, it was determined the device damage was caused by the application of improper/excessive force.Additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor material specification, found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|
|
|