C.R. BARD INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY Back to Search Results
Catalog Number 919214A
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during the pre-test, in the foley catheter, one side was bulged at a ratio of 9:1 and there was also garbage in the tray.Per sample evaluation received on 24mar2025, during sample evaluation the syringe had difficulty to deflate was found.Per sample evaluation received on 24mar2025, during sample evaluation cuffing was found.
 
Event or Problem Description
It was reported that during the pre-test, in the foley catheter, one side was bulged at a ratio of 9:1 and there was also garbage in the tray.Per sample evaluation received on 24mar2025, during sample evaluation the syringe had difficulty to deflate was found.Per sample evaluation received on 24mar2025, during sample evaluation cuffing was found.
 
Additional Manufacturer Narrative
The reported event was confirmed.Photo: received seven (7) photo samples.All photo samples showcase an overview of all-silicone temp sensing foley catheter tray and its packaging.Visual: received one (1) all-silicone temp sensing foley catheter tray.Visual inspection noted foreign material inside the tray.Using the returned syringe, the catheter balloon was inflated with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and the balloon concentricity was observed to be 100:0 as compared to attachment 1.The balloon did not passively deflate in under five (5) minutes.Using the in-house luer lock syringe, deflated balloon passively in under four (4) minutes with no leaks noted, returning 10ml of solution.However, cuffing was observed.Product labeling was reviewed for this investigation.The reported event is addressed within the labeling for this investigation.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY
Common Device Name
FOLEY TRAY
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key21815514
Report Number1018233-2025-02539
Device Sequence Number15275802
Product Code MJC
UDI-Device Identifier00801741040122
UDI-Public(01)00801741040122(11)240723(17)270428(10)MYJT3764
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number919214A
Device Lot NumberMYJT3764
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2025
Supplement Date Received by Manufacturer01/08/2026
Initial Report FDA Received Date04/10/2025
Supplement Report FDA Received Date01/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-