• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. AN INSERTER INNER SHAFT; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. AN INSERTER INNER SHAFT; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Catalog Number 602-90780
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tips of an aleutian an inserter and an aleutian an inserter inner shaft deformed intra-operatively during cage insertion, and that the cage also deformed.It was further reported that the cage remains implanted in the patient.No adverse consequence to the patient has been reported.This report captures the inner shaft.
 
Event or Problem Description
The procedure was completed with less than a three minute surgical delay.
 
Additional Manufacturer Narrative
Additional data: b5, d9, h3.Corrected data: d4, h4.H6 coding has been updated to reflect completion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AN INSERTER INNER SHAFT
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
matthew kelleher
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key21822428
Report Number3004774118-2025-00060
Device Sequence Number18466875
Product Code MAX
UDI-Device Identifier10888857157255
UDI-Public(01)10888857157255(11)220422(10)TBVB
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K150481
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number602-90780
Device Lot NumberTBVB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/12/2025
Supplement Date Received by Manufacturer07/25/2025
Initial Report FDA Received Date04/10/2025
Supplement Report FDA Received Date08/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-