• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SCF-325
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The event of infection is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: infection.
 
Event or Problem Description
Patient reported "infection".This record is for the right side.Device was explanted and replaced.
 
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further investigation with the information collected during the investigation, there is enough evidence to support that units involved in this work order were manufactured under controlled conditions in accordance with abbvie¿s procedures and were no defects/problems/issues found in the documents related to the reported event.Seal strength was successfully completed and results were acceptable.Environmental monitoring results were found to be acceptable, and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run (b)(4) is not related to any additional complaint infection record reported as of today.Considering that there is no adverse trend for this type of event, no additional actions are deemed required at this time.The issue with infection will continue to be monitored and corrective actions will be taken in the future if deemed appropriate.Given the fact that the cause of the infection cannot be specifically associated with the manufacturing process, and there is not an adverse trend for this type of event, and no er/ ncr(s) were identified during the investigation associated with this lot and the complaint, no corrective action is deemed necessary at this time.
 
Event or Problem Description
Patient reported "infection".This record is for the right side.Device was explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key21825532
Report Number9617229-2025-06072
Device Sequence Number9482310
Product Code FTR
UDI-Device Identifier10888628030077
UDI-Public(01)10888628030077(10)3475483(11)201116(17)251016
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/16/2025
Device Catalogue NumberSCF-325
Device Lot Number3475483
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2025
Supplement Date Received by Manufacturer04/14/2025
Initial Report FDA Received Date04/11/2025
Supplement Report FDA Received Date05/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-