Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during pre-test, when draining the water in the foley catheter, the sterile water did not drain even after waiting for a while.Medicon syringe was used.
|
| |
|
Event or Problem Description
|
|
It was reported that during pre-test, when draining the water in the foley catheter, the sterile water did not drain even after waiting for a while.Medicon syringe was used.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Correction: d, h.Complete initial reporter facility name: (b)(6) hospital.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during pre-test, when draining the water in the foley catheter, the sterile water did not drain even after waiting for a while.Medicon syringe was used.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon further review, bd has determined that this mdr was reported in error.There is no allegation against a bd product but on a component in the tray.This report is being submitted as additional information.The complete initial reporter's facility name is (b)(6).The reported event was confirmed- other.Received (3) photo samples.The photo sample was returned and could not be evaluated for the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), used silicone foley with syringe.Visual inspection of the sample noted no physical defects present.Using the return syringe the catheter balloon was inflated with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and inflated properly.Balloon symmetry ok.Deflated balloon with returned syringe; did not deflate passively.Inflated and deflated the balloon with (in-house) syringe; deflated passively in under 5 min: 1 min 24 sec.This complaint is against the syringe.The labeling was found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Based on the results of the investigation, no additional actions can be taken at this time.Corrections: d, e, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|