• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER PICCOLO; OCCLUDER, PATENT DUCTUS, ARTERIOSUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER PICCOLO; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number 9-PDAP-04-02-L
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2025, a 4 mm x 2 mm amplatzer piccolo occluder was chosen for a procedure using an unknown delivery system.During procedure, the device was not taking proper shape.The deformed device was never released from the delivery cable while inside the patient.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.The procedure was abandoned.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
 
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
An event of device deformity was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.Based on available information and without the device to analyze, the cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Additional Manufacturer Narrative
An event of device deformity was reported.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PICCOLO
Common Device Name
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key21833584
Report Number2135147-2025-01958
Device Sequence Number11805942
Product Code MAE
UDI-Device Identifier05415067031006
UDI-Public(01)05415067031006(17)280131(10)8921439
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P020024
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-PDAP-04-02-L
Device Catalogue Number9-PDAP-04-02-L
Device Lot Number8921439
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2025
Supplement Date Received by Manufacturer05/21/2025
02/02/2026
Initial Report FDA Received Date04/14/2025
Supplement Report FDA Received Date05/29/2025
02/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN DELIVERY SYSTEM
Patient SexUnknown
-
-