Model Number FX800T
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 03/20/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a m.Blue (#fx800t) was implanted together with a progav 2.0 and a shuntassistant 2.0 during a procedure performed on (b)(6) 2025.According to the complainant, the valve caused an underdrainage.The patient underwent a revision procedure on (b)(6) 2025.The complained product was sent to the manufacturer for investigation.No patient complications were reported as a result of the revision procedure.The release for examination has not yet been submitted.For this reason, an examination is still pending.Same incident as medwatch# 3004721439-2025-00107 and #3004721439-2025-00108.
 
Additional Manufacturer Narrative
Visual inspection: during the investigation, no significant damage of the valves was determined.Permeability test_ a permeability test has shown that the m.Blue is permeable.During the permeability test bloody liquid were flushed out.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the m.Blue operates within the accepted tolerance in both positions.Adjustment test: the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valve, no deposits were found in m.Blue.Results: based on our investigation results, we can determine no functional deviations or other abnormalities on the product.The valve is operating within its specifications.The investigation of the return shipment is complete with the creation of this report.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
The release for examination has now been submitted.And an investigation was carried out.Same incident as medwatch# 3004721439-2025-00107 and #3004721439-2025-00108.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE 0 VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam 14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key21833711
Report Number3004721439-2025-00106
Device Sequence Number4198384
Product Code JXG
UDI-Device Identifier04041906504397
UDI-Public4041906504397
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX800T
Device Catalogue NumberFX800T
Device Lot Number20075832
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2025
Supplement Date Received by Manufacturer04/24/2025
Initial Report FDA Received Date04/14/2025
Supplement Report FDA Received Date05/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexFemale
Patient Weight90 KG
-
-