RESPIRONICS, INC. REMSTAR PLUS; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number 260P
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient purchased a system one cpap device from a third-party private seller, and it has black specks coming from the device.The patient was inquiring about repairs.The patient was advised that since this device is part of the recall it could not be repaired, and they should contact the person it was purchased from.There was no report of patient harm or injury.There was no report of medical intervention.The device has not been returned for investigation.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer previously reported "the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient purchased a system one cpap device from a third-party private seller, and it has black specks coming from the device.The patient was inquiring about repairs.The patient was advised that since this device is part of the recall it could not be repaired, and they should contact the person it was purchased from.There was no report of patient harm or injury.There was no report of medical intervention.The device has not been returned for investigation." the product associated with this complaint was not returned to the manufacturer.This report is being sent to conclude that no further investigation is possible.No device has been returned.No further evaluation is possible at this time.Multiple attempts to have the device returned for evaluation and investigation were unsuccessful.In addition, capa pr 7211 documents the investigation into similar complaints, correction/removal activities have been initiated (refer to h.9), and complaint code trending is reviewed on a periodic basis to evaluate field quality performance, and as an input to risk management activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR PLUS
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key21834014
Report Number2518422-2025-104085
Device Sequence Number4288334
Product Code BZD
UDI-Device Identifier00606959028647
UDI-Public(01)00606959028647
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 12/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number260P
Device Catalogue Number260P
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2025
Supplement Date Received by Manufacturer12/04/2025
Initial Report FDA Received Date04/14/2025
Supplement Report FDA Received Date12/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/23/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-