| Catalog Number |
RR 750 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference number: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an ent procedure, when the rapid rhino was removed, it tore between the cuff and the balloon.The torn tamponade was removed from the patient with forceps.There was a surgical delay of less than 30 minutes, and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found what appears to be two rr 750: 7.5cm anterior/posterior devices.Biodebris is present, and one of the sponges is broken apart.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause of the reported event could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include tension inconsistent with normal use.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|