MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CVC SET: 3-LUMEN 7 FR X 20 CM; CATHETER PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CVC SET: 3-LUMEN 7 FR X 20 CM; CATHETER PERCUTANEOUS Back to Search Results
Model Number IPN928974
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported "when the syringe is gently pushed onto the luer connector, the puncture cannula can become hairline cracked or splinter completely.In one case, the cannula was already defective when unpacked." a new cannula was used successfully.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "unknown" at this time.Associated mdr numbers include: 3006425876-2025-00371, 3006425876-2025-00370, and 3006425876-2025-00369.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return; however, the customer provided two photos for analysis.Based on visual analysis of the customer photos, the reported event is confirmed.A device history record review was performed and no relevant findings were identified.The ifu provided with the kit contains the following warning for users: "some disinfectants used at catheter insertion site contain solvents which can weaken the catheter material.Alcohol, acetone, and polyethylene glycol can weaken the structure of polyurethane materials.These agents may also weaken the adhesive bond between catheter stabilization device and skin.Do not use acetone on catheter surface.Do not use alcohol to soak catheter surface or allow alcohol to dwell in a catheter lumen to restore catheter patency or as an infection prevention measure.Do not use polyethylene glycol containing ointments at insertion site.Take care when infusing drugs with a high concentration of alcohol.Allow insertion site to dry completely prior to skin puncture and before applying dressing.Do not allow kit components to come into contact with alcohol." the customer report of a cracked/separated needle hub was confirmed through visual inspection of the returned customer photo.The customer photo revealed that the needle hub was cracked and separated.The failure mode identified is consistent with the failure mode investigated per a previously opened capa.Based on the capa investigation, the root cause of this complaint is design related.Teleflex has identified that the needle hub material is susceptible to cracking when placed under stress (i.E.Pressed onto a tapered luer fitting, sideloaded as the clinician attempts to locate a vessel, etc.) in the presence of liquid alcohol-based disinfectants.The capa was implemented using a new alcohol-resistant material to manufacture the needle hub.The needle hub from this complaint was confirmed to be made from the old needle hub material.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported "when the syringe is gently pushed onto the luer connector, the puncture cannula can become hairline cracked or splinter completely.In one case, the cannula was already defective when unpacked." a new cannula was used successfully.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "unknown" at this time.Associated mdr numbers include: 3006425876-2025-00371, 3006425876-2025-00370, and 3006425876-2025-00369.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 3-LUMEN 7 FR X 20 CM
Common Device Name
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key21838465
Report Number3006425876-2025-00371
Device Sequence Number7963187
Product Code DQY
UDI-Device Identifier40801902127714
UDI-Public(01)40801902127714(11)240925(17)290731(10)71F24H0852(21)
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K862056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN928974
Device Catalogue NumberCS-15703-E
Device Lot Number71F24H0852
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2025
Supplement Date Received by Manufacturer05/14/2025
Initial Report FDA Received Date04/15/2025
Supplement Report FDA Received Date05/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-