Model Number M00583100
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a truetome jag 44 was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2025.During the procedure, the truetome was used to cut the papilla after the intubation.The power supply was connected.Upon energizing, the cutting wire of the truetome jag 44 broke at the rear end.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h11: (investigation result) the returned truetome jag 44 was analyzed, and a visual evaluation noted that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Additionally, the working length was twisted.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Based on the condition of the device, the wire break could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Additionally, the working length twisted could be caused after multiple attempts to rotate handle of the device or during the introduction of the device into the scope.Based on all gathered information, the most probable root cause is adverse event related to procedure.
 
Event or Problem Description
It was reported to boston scientific corporation that a truetome jag 44 was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2025.During the procedure, the truetome was used to cut the papilla after the intubation.The power supply was connected.Upon energizing, the cutting wire of the truetome jag 44 broke at the rear end.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETOME JAG 44
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
arlene acevedo gutierrez
300 boston scientific way
marlborough, MA 01752
7877968283
MDR Report Key21841000
Report Number3005099803-2025-01601
Device Sequence Number20798669
Product Code KNS
UDI-Device Identifier08714729858645
UDI-Public(01)08714729858645(17)260723(10)34493522
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K122203
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/23/2026
Device Model NumberM00583100
Device Catalogue NumberM00583100
Device Lot Number0034493522
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2025
Supplement Date Received by Manufacturer05/14/2025
Initial Report FDA Received Date04/15/2025
Supplement Report FDA Received Date06/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
Patient Weight53 KG
-
-