• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XXL 360 12MM X 45CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XXL 360 12MM X 45CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0036181240
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the coil (subject device) got detached within the microcatheter.The subject device was replaced, and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.It was reported that the coil was detached inside the microcatheter while advancing and the coil was removed from the patient, together with the microcatheter.An assignable cause of undeterminable will be assigned to this complaint as the product was not returned and review and analysis of all available information fails to indicate an assignable cause or probable assignable cause.
 
Event or Problem Description
It was reported that the coil (subject device) got detached within the microcatheter.The subject device was replaced, and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XXL 360 12MM X 45CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key21841153
Report Number3008881809-2025-00207
Device Sequence Number8817385
Product Code HCG
UDI-Device Identifier07613327128307
UDI-Public(01)07613327128307(17)270729(10)24990669
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K242243
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberM0036181240
Device Lot Number24990669
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/24/2025
Supplement Date Received by Manufacturer05/13/2025
Initial Report FDA Received Date04/15/2025
Supplement Report FDA Received Date06/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BREAKTHROUGH MICROCATHETER (NON-STRYKER)
Patient SexUnknown
-
-