|
Correction: h3.Yes h6.Investigation findings: 3243 investigation conclusions: 61 device evaluation: visual inspection confirms that the hexalobe tip is fractured.The root cause is likely due to wear as a result of improper loading before torque delivery through the tip.Review of device history records showed no manufacturing or processing related irregularities.Labeling review: warnings: risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury.Possible adverse effects: initial or delayed loosening, disassembly, bending, dislocation and/or breakage of device components.
|