| Model Number |
N164 |
| Medical Device Problem Codes |
High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a retrospective review of device data it was identified that during a shock delivery, this system recorded a code 1005 indicative of high, out of range shock impedance measurements greater than 145 ohms.No other information was provided.No additional adverse patient effects were reported.This device was explanted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a retrospective review of device data it was identified that during a shock delivery, this system recorded a code 1005 indicative of high, out-of-range shock impedance measurements greater than 145 ohms.No other information was provided.No additional adverse patient effects were reported.This device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of device displays error message was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.
|
| |
|
Search Alerts/Recalls
|