• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA? CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA? CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER Back to Search Results
Model Number N164
Medical Device Problem Codes High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a retrospective review of device data it was identified that during a shock delivery, this system recorded a code 1005 indicative of high, out of range shock impedance measurements greater than 145 ohms.No other information was provided.No additional adverse patient effects were reported.This device was explanted.
 
Event or Problem Description
It was reported that during a retrospective review of device data it was identified that during a shock delivery, this system recorded a code 1005 indicative of high, out-of-range shock impedance measurements greater than 145 ohms.No other information was provided.No additional adverse patient effects were reported.This device was explanted.
 
Additional Manufacturer Narrative
A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of device displays error message was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA? CRT-D
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21853499
Report Number2124215-2025-24148
Device Sequence Number8699626
Product Code LWP
UDI-Device Identifier00802526481031
UDI-Public(01)00802526481031(17)161008
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/08/2016
Device Model NumberN164
Device Catalogue NumberN164
Device Lot Number106322
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2025
Supplement Date Received by Manufacturer05/29/2026
Initial Report FDA Received Date04/17/2025
Supplement Report FDA Received Date06/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/17/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-