| Catalog Number |
SSM-275 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
03/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported capsular contracture, baker grade iii.This record is for the left side.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted the event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
The device has been explanted and replaced.Prednisone taper and singulair were used as treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: a5, a6, b5, b7, d6b, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture baker grade iii".This record is for the left side.Device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d.9, h.3, h.6.Device evaluation: based on the product analysis performed, the assessments of the complaint codes are: ¿ capsular contracture: unable to observe.As per the investigation procedure, deformation were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|