• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP,TSV,4.1,13,MTX,MG; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP,TSV,4.1,13,MTX,MG; DENTAL IMPLANT Back to Search Results
Catalog Number TSVT4B13
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight unknown / not provided.
 
Event or Problem Description
It was reported crown came off implant.Attempted to place healing collar on implant until dentist could replace crown, but the hex of abutment was broken.Attempted to remove broken hex but was unable to.Therefore implant was removed and replaced with 4.1x13 and bone graft.
 
Additional Manufacturer Narrative
Zimvie received one (1) portion / fragment of the unknown abutment for evaluation.Visual evaluation of the as returned devices identified the fractured abutment hex stuck inside the implant.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Dhr, sterilization, and complaint history review could not be performed, as the subject lot number associated with the unknown abutment is not available.Zimvie quality management system (qms) has controls in place to prevent the distribution of non-conforming product and ensure the product is within specifications.The customer did not submit images for the reported event.Based on the investigation and risk management file review as per rmf (b)(4) rev.5, the most likely root causes determined from the investigation are clinician error, or patient factors.Therefore, based on the available information, a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.G3: date received by manufacturer.G6: checked "follow-up".H2: checked follow-up type.H3: changed "no" to "yes".H6: entered evaluation codes.H10: added manufacturer narrative.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,4.1,13,MTX,MG
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key21855629
Report Number0002023141-2025-00995
Device Sequence Number7841482
Product Code DZE
UDI-Device Identifier00889024019843
UDI-Public(01)00889024019843(17)230530(10)1216951
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K101977
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/30/2023
Device Catalogue NumberTSVT4B13
Device Lot Number1216951
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/17/2025
Supplement Report FDA Received Date07/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/30/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-