MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 CLOSED WOUND SUCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 CLOSED WOUND SUCTION Back to Search Results
Catalog Number 0043650
Medical Device Problem Code Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the following two problems had occurred in the spring evacuator.Spring evacuator expands even though it was clamped.Leaks were occurring from the crimped part at the end.Per follow up information received via ibc on 02apr2025, customer confirmed that patient involvement.
 
Additional Manufacturer Narrative
The reported event was inconclusive.The conditions of the sample did not allow further evaluation.Visual evaluation of the returned seven-photo sample noted one opened (without original packaging), used 3-spring closed wound suction evacuator.Visual inspection of the first photo sample noted an overview of the 3-spring, drain and syringe.The second photo shows the front view of the 3-spring evacuator.The third and fourth photo shows the side view of the 3-spring evacuator.The fifth and sixth photo shows the overview of the drain.The seventh photo shows the product packaging with the product information.No actions can be taken at this time since a root cause was not identified.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and states the following: indications: wound drains are used to remove exudates from wound sites.Warnings: an effective closed suction drain system requires maintenance of the system to preserve patency.The drain must not be allowed to occlude nor the reservoir to completely fill; and reservoir suction must be maintained in order for the system to function properly.Verify that the system is functioning properly.If the system is not maintained properly, surgical complications, including hematomas, may result.In the event of occlusion of the drain, all wound drainage via the drain ceases.Careful attention to the drain will minimize the possibility of this problem.If occlusion does occur, the drain can be aspirated by connecting suction to the reservoir outlet or temporarily disconnecting the drain from the reservoir and applying suction directly to the drain.If an air-tight seal between the drain and the skin where the drain emerges is not achieved, the air leak must be rectified or the system must be converted to open drainage.An airtight seal between all system components (drain, adaptor and reservoir) is necessary for proper system function.Leaving the soft silicone elastomer drain implanted for any period of time so as to cause tissue ingrowth around the drain can interfere with easy removal and may effect the performance of the drain.The surgeon should monitor the patient¿s rate of wound healing.Evacuators should be used in cardio-thoracic surgery only after the lung is fully expanded and all air leaks have sealed.Drain perforations or channels must lie within the wound or cavity to be drained, otherwise inadequate drainage may result.To avoid the possibility of drain damage or breakage: avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked for free motion during closure to minimize the possibility of breakage.Drain removal should be done gently by hand.Drains should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Surgical removal may be necessary if drain is difficult to remove or breaks.Complications complications which may result from the use of this suction drainage system include the risks associated with methods utilized in the surgical procedure, as well as the patients degree of intolerance to any foreign object in the body.The advantages of wound drainage, particularly closed system drainage, are lost if an airtight seal between the drain and the skin where the drain emerges is not achieved, or if the drain is allowed to become occluded or if the reservoir is not activated properly, doesn¿t function properly or is not monitored.Evacuators should be emptied and re-activated when required per hospital protocol.Correction- d, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the following two problems had occurred in the spring evacuator.Spring evacuator expands even though it was clamped.Leaks were occurring from the crimped part at the end.Per follow up information received via ibc on 02apr2025, customer confirmed that patient involvement.
 
Additional Manufacturer Narrative
The reported event is unconfirmed.No root cause could be found because the reported event was unconfirmed.Visual evaluation noted received 1 evacuator with tubing and syringe in opened packaging.Verified material number 0043650 and batch number ngjs4281.No physical defects present.Further testing required.Visual evaluation of the returned seven-photo sample noted one opened (without original packaging), used 3-spring closed wound suction evacuator.Visual inspection of the first photo sample noted an overview of the 3-spring, drain and syringe.The second photo shows the front view of the 3-spring evacuator.The third and fourth photo shows the side view of the 3-spring evacuator.The fifth and sixth photo shows the overview of the drain.The seventh photo shows the product packaging with the product information.Dhr review is not required as the reported event is unconfirmed however the dhr review was completed prior to sample evaluation update.Labelling review is not required as the reported event is unconfirmed.Correction: h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the following two problems had occurred in the spring evacuator.Spring evacuator expands even though it was clamped.Leaks were occurring from the crimped part at the end.Per follow up information received via ibc on 02apr2025, customer confirmed that patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLOSED WOUND SUCTION
Common Device Name
CLOSED WOUND SUCTION
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key21855771
Report Number1018233-2025-02734
Device Sequence Number14623802
Product Code GCY
UDI-Device Identifier00801741049385
UDI-Public(01)00801741049385(11)240706(17)290430(10)NGJS4281
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0043650
Device Lot NumberNGJS4281
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2025
Supplement Date Received by Manufacturer05/07/2025
07/03/2025
Initial Report FDA Received Date04/17/2025
Supplement Report FDA Received Date05/09/2025
07/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-