• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO PHILIPPINES CORPORATION SURGUARD3 SAFETY HYPODERMIC NEEDLE (MCKESSON PRIVATE LABEL); NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO PHILIPPINES CORPORATION SURGUARD3 SAFETY HYPODERMIC NEEDLE (MCKESSON PRIVATE LABEL); NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number 102-N251S3
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 02/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D6a: implanted date: device was not implanted.D6b: explanted date: device was not explanted.E3: occupation: rn administrator qa/pi.G4: pma/510(k): not available.The details of the sample's actual condition were unable to be determined as it was not available for evaluation.Retention samples were visually inspected and found to be free of any damage or irregularities on the sheath and collar.Sheath activation, sheath deactivation, and component fit tests were performed on the samples, and all passed.There was no issued nonconformity report related to human error during lot production.There are no nonconformities in the assembled needle lot supplied during production.During the collar assembly process, the hub fit is loosened from the protector before placing the needle into the collar.Prior to proceeding to the next process, the hub, cannula, and collar are pressed into the protector wherein a photoelectric sensor detects if the height of the protector is correct.Our safety needle features a hinged safety sheath attached to the needle hub.The safety sheath contains two locking mechanisms, the sheath tooth-cannula, and sheath wings collar, which are simultaneously activated when manually pressed over the needle immediately after use and just before disposal to minimize the possibility of sharps injury.The collar and sheath undergo an initial product inspection check (ipic) before mass production, during raw material change, mold maintenance, plant shutdown, cavity closure, and mold/product type change.Any molding defects that may affect product function are part of the inspection item.We also have a visual in-process and product confirmation inspection to ensure that the collar and sheath are in good condition.Visual in-process inspections are conducted during the manufacturing process to detect abnormalities on the product that may cause issues during sheath activation.The critical locking part of the sg3 product is checked for defects such as short shot, sheath collar fitting, and over or under insertion of the collar into the hub.We have functional tests conducted to confirm that our products are free of defects that would lead to difficulty or failure of device activation.We perform safety sheath activation during sg assembly inspection and outgoing inspection of finished products where the cannula should be captured under the sheath's tooth.During the component fit test, we also check that the sheath lugs are securely attached to the collar ear.Before shipment, qc conducts outgoing inspections to check the overall condition of the products.The product's instructions for use (ifu) provide detailed instructions for using the sg3 safety needle device, including warnings to prevent needle sticks.The collar and sheath are manufactured in-house with validated tpc molding machines.There were no nonconformance reports issued for the supplied molded part lots.From the previous three fiscal years to the present, we received a total of five (5) complaints for the specific item code.The cause of these complaints was not identified as being related to our product or the manufacturing process.Retention samples of 25g needles were simulated.The samples were activated using three methods: finger, thumb, and hard surface activation.All of the samples were successfully activated, and the cannula was captured under the sheath tooth.There were no difficulties encountered during the activation.Based on the investigation, we cannot identify the cause of the problem.Limited information was provided related to the complaint.A review of the product's lot history file revealed no related issues that would prevent the safety sheath from activating, resulting in a needlestick injury to the user.We have routine inspections to check the condition of the products.The components that are critical to the product's safety activation are checked to ensure that no defects occur that could result in sheath activation issues.We recommend following the instructions for use (ifu) for proper use of the sg3 needle, which includes caution, precautions, and warnings to avoid problems during sheath activation.Terumo medical corporation (tmc) (importer) registration no.2243441 is submitting this report on behalf of terumo (philippines) corporation (manufacturer) registration no.3003902955.
 
Event or Problem Description
The user facility reported that the retraction mechanism of the device is jamming, which resulted in one needlestick.The customer discarded the remainder of the box, so no sample is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGUARD3 SAFETY HYPODERMIC NEEDLE (MCKESSON PRIVATE LABEL)
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan laguna, 4024
RP  4024
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan laguna, 4024
RP   4024
Manufacturer Contact
stephanie handy
950 elkton blvd.
elkton, MD 21921
9499890491
MDR Report Key21856277
Report Number3003902955-2025-00019
Device Sequence Number4721800
Product Code FMI
UDI-Device Identifier10612479246390
UDI-Public(01)10612479246390(17)290430(10)240507E
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number102-N251S3
Device Catalogue NumberN/A
Device Lot Number240507E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2025
Initial Report FDA Received Date04/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-