BOSTON SCIENTIFIC CORPORATION RUBICON 35; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 39540
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of the wire punctured through the catheter.The target lesion was located in the superficial femoral artery.A 135cm rubicon 35 catheter guide was used for leg angiography procedure and was placed over the wire.After the wire was withdrawn, the zipwire hydrophilic guide wire was inserted into the rubicon catheter, but advancing was difficult.It was noted that the wire punctured through the rubicon, causing a hole.Both devices were removed, and the center tip of the wire was found protruding from the outer layer of the wire.A new guidewire and rubicon were opened and completed the procedure.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBICON 35
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key21857274
Report Number2124215-2025-23259
Device Sequence Number4719935
Product Code DQY
UDI-Device Identifier08714729832072
UDI-Public(01)08714729832072(17)271211(10)35474150
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K122394
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number39540
Device Catalogue Number39540
Device Lot Number0035474150
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2025
Initial Report FDA Received Date04/17/2025
Date Device Manufactured12/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-