H3: product analysis: report not confirmed.Evaluation could not reproduce the reported malfunction of clearview is without drilling tip.However, it was noted that bur tip on clear view tip was fractured off.This finding may have contributed to the user¿s experience.It was also noted that the fracture occurred under rotational load, which indicates that it failed during use.This regulatory report is being submitted due to a retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|