The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record could not be completed.The part number/lot number combination needed for device identification remains unknown.The user facility was unable to provide this information.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident could not be completed.No medical records nor medical imaging relevant to the reported adverse event/incident was received by endologix.The user facility was unable to provide this information.Due to the absence of medical records and medical imaging; device, use, procedure, and/or anatomy relatedness to this adverse event/incident could not be evaluated.It was reported that the patient was discharged on postoperative day five.The patient status was not reported.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.
|
|
The patient was initially implanted with an afx2 bifurcated stent graft (bsg) and an afx vela suprarenal aortic extension to treat an abdominal aortic aneurysm (aaa) on (b)(6) 2019.It was reported that four (4) years post initial procedure the contrast-enhanced computed tomography (ct) showed that there was aneurysm enlargement, enlarging from 45.0 mm to 57.3 mm and a type iiia endoleak (implant separation) between the afx2 bsg and afx vela suprarenal aortic extension.On (b)(6) 2024, an afx vela suprarenal aortic extension was implanted overlapping the afx2 bsg and afx vela suprarenal aortic extension by 5 cm each.The intraoperative contrast confirmed that the type iiia endoleak was resolved.It was reported that the patient was discharged on postoperative day five.The patient status was not reported.
|