| Catalog Number |
8065982465 |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A nurse reported that an ophthalmic blade was found to be dull and it was not cutting.The another blade was opened to completed the procedure by making the necessary cut.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.A sample is available that has not yet been received at manufacturing for evaluation.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The sample was visually inspected and found to be nonconforming with damaged cutting edge and nick in the blade.Penetration testing could not be performed due to the damage of the sample.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The exact root cause could not be determined from the investigation performed.The returned sample is visually non-conforming with the damaged cutting edge and nick in the blade.The damage to the returned sample is consistent with damage that can occur when the blade contacts a hard surface such as the protective blade tray when product is improperly removed or inserted after use, or improper handling, or contact with another instrument during surgery or set-up.The damage seen on the returned complaint sample could have contributed to the customers reported issue of dull blade.The exact root cause for the damaged knife sample is unknown, therefore specific action with regards to this complaint cannot be taken.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Any nonconformance, such as the and damaged cutting edge exhibited on the returned opened sample, is removed from the lot and scrapped.Functional penetration testing is performed and monitored during the finishing process to ensure the sharpness of the product.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|