• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CLEARCUT KNIVES; KNIFE, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CLEARCUT KNIVES; KNIFE, OPHTHALMIC Back to Search Results
Catalog Number 8065982465
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
A nurse reported that an ophthalmic blade was found to be dull and it was not cutting.The another blade was opened to completed the procedure by making the necessary cut.There was no patient harm.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.A sample is available that has not yet been received at manufacturing for evaluation.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The sample was visually inspected and found to be nonconforming with damaged cutting edge and nick in the blade.Penetration testing could not be performed due to the damage of the sample.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The exact root cause could not be determined from the investigation performed.The returned sample is visually non-conforming with the damaged cutting edge and nick in the blade.The damage to the returned sample is consistent with damage that can occur when the blade contacts a hard surface such as the protective blade tray when product is improperly removed or inserted after use, or improper handling, or contact with another instrument during surgery or set-up.The damage seen on the returned complaint sample could have contributed to the customers reported issue of dull blade.The exact root cause for the damaged knife sample is unknown, therefore specific action with regards to this complaint cannot be taken.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Any nonconformance, such as the and damaged cutting edge exhibited on the returned opened sample, is removed from the lot and scrapped.Functional penetration testing is performed and monitored during the finishing process to ensure the sharpness of the product.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARCUT KNIVES
Common Device Name
KNIFE, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21857552
Report Number2523835-2025-00484
Device Sequence Number15449096
Product Code HNN
UDI-Device Identifier00380659824654
UDI-Public(01)00380659824654(17)291130(11)241204(10)1762HJ
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065982465
Device Lot Number1762HJ
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/08/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/26/2025
Supplement Date Received by Manufacturer06/24/2025
Initial Report FDA Received Date04/17/2025
Supplement Report FDA Received Date07/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CUSTOM-PAK SURGICAL; CUSTOM-PAK SURGICAL PROCEDURE PACK
Patient SexUnknown
-
-