| Model Number |
324911 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This medwatch submission is both an initial and supplemental filing.Investigation of the results can be seen below: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformance's were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
| |
|
Event or Problem Description
|
|
Spouse of consumer reported that the plunger rod will not depress when doing injection.He stated that he was able to draw the insulin from the vial.Consumer does not re-use.Lot#: 1350910, catalog#: 324911, date of event: unknown, samples: discarded.
|
| |
|