| Model Number |
EM200 |
| Medical Device Problem Codes |
Overheating of Device (1437); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Repair request initiated for device with the report of device overheating and noisy.It was reported that there was no patient invol vement.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation could not be performed because motor run rough.It was noted that top of motor and bottom of collet corroded.This finding may have contributed to the user's experience.It was also noted motor noisy.This regulatory report is being submitted due to a retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|