| Model Number |
G34266 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an endoscopic procedure, the physician used a cook acrobat¿ calibrated tip wire guide.It was reported that the covering peels/detaches from wire guide during stent implantation, replacement of wire guide [loss of wire guide access] and continuation of cannulation and implantation [was required].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: the product said to be involved was returned in a white plastic bag.A lot number was not provided with the returned device.Our laboratory evaluation of the product said to be involved confirmed the report.There is wire guide coating damage near the distal end.A section of the coating has split and peeled exposing the core wire approximately 14.3cm to 26.9cm from the distal end.The coating was partially socked 13.7cm to 14.3cm from the distal tip and was able to be unsocked.No portion of the coating appears to be missing.The straight split lines with a shallow cut in the coating continuing from the split and peeled section was observed under magnification, evidence of potential scoring damage contributing to the failure.A lab meeting was held with production leadership and manufacturing engineering on 14apr2025.A visual inspection of the coating damage confirmed evidence of a scoring defect.This is the most likely cause for the coating damage and is considered operator related.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the returned device confirmed the report.The coating at the distal end was damaged in manufacturing during the scoring process.This occurred due to operator error.Production management and the department team leads were notified of this occurrence.Additionally, a notification of operator related complaint form was provided to production management to make them aware of an operator related complaint.However, retraining was not performed in response to this occurrence as the manufacturing process has been revised since the production date of this device, removing the scoring process from the manufacturing process.A corrective action (capa) was initiated to investigate device failure related to coating damage caused by the scoring fixture.This device is within the scope of the capa.Prior to distribution, all acrobat¿ calibrated tip wire guides are subjected to a visual inspection to ensure device integrity.Corrective action: a corrective action (capa) was initiated to investigate device failure due to coating damage during the scoring process.The product said to be involved is included in the scope of the corrective action.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends.
|
| |
|
Search Alerts/Recalls
|