Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The bactiseal catheter (id 821749) was returned for evaluation.Device history record (dhr) - the product code 821749 with lot 7365500, conformed to the specifications when released to stock.Failure analysis - the catheter was visually inspected.The leak test was performed; a leak was noted.The complaint was confirmed.Root cause analysis - a possible root cause for the issue reported by the customer, could be due to a sharp or pointed object coming into contact with the catheter, as noted in the ifu silicone has a low cut/tear resistance.
|