• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSUREFIX LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSUREFIX LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 5568-53
Medical Device Problem Codes Over-Sensing (1438); Under-Sensing (1661)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the right atrial (ra) lead exhibited undersensing at times and far field r-wave (ffrw) oversensing resulting in inappropriate mode switching.It was also reported that there were possible high left ventricular (lv) lead thresholds and high right ventricular (rv) lead thresholds noted.Additionally, the cardiac resynchronization therapy defibrillator (crt-d) inappropriately withheld therapy for ventricular tachycardia (vt) with 1:1 retrograde detected as supra ventricular tachycardia (svt) via the device feature that uses pattern and rate analysis to discriminate between svt and true ventricular tachyarrhythmias.It was also previously observed that the crt-d inappropriately withheld therapy for vt detected as svt via the device feature designed to withhold inappropriate ventricular detection if the rhythm displays characteristics of an svt origin.The ra, rv and lv leads and device remain in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: dtpa2q1 crt-d implanted: (b)(6) 2024.429888 lead implanted: (b)(6) 2014 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the right atrial (ra) lead exhibited undersensing at times and far field r-wave (ffrw) oversensing resulting in inappropriate mode switching.It was also reported that there were possible high left ventricular (lv) lead thresholds and high right ventricular (rv) lead thresholds noted.Additionally, the cardiac resynchronization therapy defibrillator (crt-d) inappropriately withheld therapy for ventricular tachycardia (vt) with 1:1 retrograde detected as supra ventricular tachycardia (svt) via the device feature that uses pattern and rate analysis to discriminate between svt and true ventricular tachyarrhythmias.It was also previously observed that the crt-d inappropriately withheld therapy for vt detected as svt via the device feature designed to withhold inappropriate ventricular detection if the rhythm displays characteristics of an svt origin.The ra, rv and lv leads and device remain in use.No further patient complications have been reported as a result of this event.It was further reported that the patient was seen in clinic and the lv lead and rv lead thresholds were within normal limits.The ra lead and crt-d were reprogrammed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX LEAD
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key21859026
Report Number2649622-2025-10477
Device Sequence Number7982009
Product Code DTB
UDI-Device Identifier00885074248592
UDI-Public00885074248592
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/19/2009
Device Model Number5568-53
Device Catalogue Number5568-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2025
Supplement Date Received by Manufacturer04/18/2025
Initial Report FDA Received Date04/17/2025
Supplement Report FDA Received Date05/01/2025
Date Device Manufactured11/28/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
694765 LEAD
Outcome Attributed to Adverse Event Required Intervention; Life Threatening;
Patient Age57 YR
Patient SexMale
Patient Weight103 KG
-
-