MAUDE Adverse Event Report: ARTHREX, INC. 5.5MM BC CORKSCREW FT W/ SUTURETAPE; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 5.5MM BC CORKSCREW FT W/ SUTURETAPE; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number 5.5MM BC CORKSCREW FT W/ SUTURETAPE
Medical Device Problem Codes Positioning Failure (1158); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
It was reported that during a rotator cuff surgery the device could not be screwed into the bone.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5MM BC CORKSCREW FT W/ SUTURETAPE
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21859118
Report Number1220246-2025-01523
Device Sequence Number7888874
Product Code MAI
UDI-Device Identifier00888867023390
UDI-Public00888867023390
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K082810
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5.5MM BC CORKSCREW FT W/ SUTURETAPE
Device Catalogue NumberAR-1927BCT
Device Lot Number15295770
Was Device Available for Evaluation? No
Distributor Facility Aware Date04/01/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2025
Supplement Date Received by Manufacturer08/11/2025
Initial Report FDA Received Date04/17/2025
Supplement Report FDA Received Date08/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-