| Model Number |
VTICM5 12.6 (-13.0/1.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens, is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced lens rotation not associated with low vaulting.The lens was exchanged with a same length but different axis lens and this resolved the problem.The cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: lens was returned in a micro-centrifuge vial, dry with residue/debris on the product.Visual inspection found no visible damage to the lens.Dimensional inspection found the lens within specifications.Claim#: (b)(4).
|
| |
|