• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 INTRAOSSEOUS, INTRAOSSEOUS NEEDLE KITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 INTRAOSSEOUS, INTRAOSSEOUS NEEDLE KITS Back to Search Results
Model Number N/A
Medical Device Problem Codes Material Deformation (2976); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2025
Type of Reportable Event Death
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The customer reported the patient passed away but did not provide the patient's date of death.The date in this report is the date bd became aware of the patient's death.The manufacturer has received the sample and it will be evaluated.A supplemental will be submitted with evaluation results.
 
Event or Problem Description
It was reported by the customer, "during a cardiac arrest last night, bent needle that made it impossible to place an io device.Paramedic was unable to place a io during a cardiac arrest due to needles et bending during placement.Paramedic did place an iv, however, patient expired.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of an intraosseous (io) needle bending during attempted insertion was confirmed and the cause appeared to be use-related.The product returned for evaluation was one 25 mm intraosseous (io) needle.Usage residues were observed on the sample.The obturator was inlaid through the needle shaft.A bend was observed in the needle shaft near the luer adapter.The needle tip exhibited deformation.Microscopic inspection of the needle tip confirmed deformation.The needle tip was twisted and peeled away from the obturator tip.The obturator tip was folded toward the needle bevel.The bend in the needle shaft was consistent with sheer, or lateral stress placed on the needle shaft.The tip deformation was consistent with excessive force placed on the needle tip while the needle was rotating.The needle bend further suggested that force was applied while the needle was not placed at a 90 degree angle to the patient¿s skin.The product instructions for use (ifu) inform users to use moderate steady pressure when inserting the needle into the bone at a 90 degree angle to the skin.Positioning the needle in this fashion minimizes the amount of lateral pressure, reducing the risk of needle deformation/damage.The observed needle deformation appeared to be the result of insertion techniques.It is unknown to what extent initial needle placement failure ultimately contributed to declining patient conditions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAOSSEOUS, INTRAOSSEOUS NEEDLE KITS
Common Device Name
INTRAOSSEOUS NEEDLE KIT
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
kayla bevins
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key21860542
Report Number3006260740-2025-02429
Device Sequence Number7842117
Product Code FMI
UDI-Device Identifier00801741163678
UDI-Public(01)00801741163678
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K203193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberD015251NK
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2025
Supplement Date Received by Manufacturer06/17/2025
Initial Report FDA Received Date04/17/2025
Supplement Report FDA Received Date07/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-